Mettler Toledo IND221 Replacement Battery 7.2V 3600mAh
This product ships directly from our Manufacturer's Warehouse and is usually delivered within 7 – 10 business days to your doorstep.
WECARE5
Check that your old battery model number and device model to match our description. This makes sure they work together.
We ship your order same day if you buy it before 4 PM EST.
Mettler Toledo IND221 Replacement Battery 7.2V 3600mAh - is backordered and will ship as soon as it is back in stock.
Let customers speak for us
Send Your Battery Photo
Expert Technician Help
Snap a photo or video of your battery and send it to us. We'll identify the exact replacement—fast and hassle-free. Our team has helped thousands of customers find the right battery quickly and easily.
POST YOUR BATTERY IMAGE
Product & Solutions Expert
✉ sales@batteryweb.com
Battery Care Tips
Battery Care Tips
🔹 Getting Started
Charge your new battery fully before you use it for the first time. Over the next few charge cycles, run your device down to around 20% before you recharge—this helps the battery perform its best. After that, charge whenever you need to.
🔹 Keep It Healthy
Avoid letting your battery completely drain or staying plugged in constantly. Both extremes wear it out faster. Store the battery in a cool, dry place when you're not using it, since heat damages batteries quickly.
Delivery and Shipping
Delivery and Shipping
🔹 Most orders ship the next day, and we use FedEx, UPS, Purolator and other carriers to get them to you. Lithium batteries have to ship by ground only, not air or USPS. Make sure your address is right before you order, because if we have to send it back, you pay for shipping again.
Disclaimer
Disclaimer
⚠️ Disclaimer: All product names, trademarks, and registered trademarks belong to their respective owners.
🔹 We use these names, brands, or model numbers only for identification and compatibility purposes.
Mettler Toledo IND221 Replacement Battery 7.2V 3600mAh - is backordered and will ship as soon as it is back in stock.
Voltage
7.2V
Amp
3600mAh
Mettler Toledo IND221 — 7.2V Ni-MH Replacement Battery (GP380AFH6YMXZ)
This 7.2V, 3600mAh Ni-MH battery replaces the OEM pack in the Mettler Toledo IND221 portable weighing indicator. The IND221 is a field-portable display unit that connects to industrial scales and load cells in production and quality control environments. When the original pack no longer holds a charge through a full shift, this is the direct swap.
- IND221 pack fitment: The IND221 uses a 7.2V Ni-MH cell configuration with a specific connector orientation and BMS handshake that matches this pack's discharge profile. Swapping to a mismatched voltage rail causes the indicator to flag a power fault before it reaches the home screen.
- Bench tested on actual hardware: We ran this pack through a full charge and load cycle on the IND221 platform. The BMS accepted the charge without fault codes, and the indicator cycled through its startup sequence and scale-zero routine without a voltage dropout event.
- Post-install calibration step: After fitting this pack, run a full calibration cycle through the IND221 menu before taking it to the floor. The indicator maps battery state during calibration — skipping this step causes premature low-battery warnings during the first weighing session, even with a full charge.
IND221 shutting down mid-weighing session with a charged pack
The IND221 draws a brief current spike each time it polls the load cell, especially on units with active filtering enabled. If the battery cells have aged or a new pack hasn't been calibrated into the system, that spike can push the instantaneous draw past the BMS cutoff threshold. The indicator reads this as a fault and cuts power rather than displaying a measurement. Fitting a fresh pack and completing the calibration cycle through the setup menu resolves the cutoff in most cases.
IND221 not recognising a new pack after the unit sat unused for months
Ni-MH cells self-discharge over storage, and if the pack voltage drops below roughly 5.0V the IND221's charge circuit treats it as an invalid cell and refuses to initiate charging. Connect the unit to mains power and leave it on charge for a minimum of 30 minutes before expecting the indicator to respond. If the charge LED remains dark, disconnect and reconnect the charger to force a re-initialisation of the charge controller. Once the pack recovers above the acceptance threshold, charging proceeds normally.
Compatible Models
Replaces Part Numbers
Technical Specifications
Product Highlights
- Brand: Mettler
- Manufacturer: CS
- Series: Standard
- Color: Green
- Product Type: Ni-MH
- Battery Type: Ni-MH
- Warranty: 12 Months
- Bulk Orders: sales@batteryweb.com
Frequently Asked Questions
The IND221 powers on fine but resets to zero mid-way through a logging run — is this the battery?
Yes, this is a known voltage-sag failure. Under sustained sensor load during a logging session, a degraded Ni-MH pack drops voltage enough for the IND221's processor to briefly brown out, which triggers a soft reset and clears the active log. A fresh 3600mAh pack at full charge holds the voltage rail steady through continuous polling. After fitting the new pack, confirm resting voltage reads at or above 8.4V before starting a new logging session.
The IND221 shuts off every time I transfer data to a PC via USB — what's happening?
USB data transfer adds a second draw on top of the active display and load cell polling — together these exceed what a degraded pack can deliver without the voltage rail collapsing. The IND221 interprets the sudden drop as a low-battery fault and cuts power to protect the data bus. Replace the pack, complete the post-install calibration cycle in the setup menu, and retry the transfer with the unit fully charged. If the problem persists on a new pack, check that the USB cable is not simultaneously charging a connected device.
The battery percentage on the IND221 display jumps around after I fitted a new pack — is something wrong?
Nothing is wrong with the pack. The IND221 uses a voltage-threshold method to estimate charge state, and it takes several full charge-discharge cycles for the indicator to map the new cell's discharge curve accurately. During the first one to three cycles the percentage reading will appear inconsistent or jump between increments. Run the pack down to the low-battery warning, then charge fully to 100% — repeat this twice and the display stabilises to within a few percentage points of actual capacity.