Fluke 700 Calibrator Replacement Battery 7.2V 2500mAh
Check that your old battery model number and device model to match our description. This makes sure they work together.
We ship your order same day if you buy it before 4 PM EST.
Fluke 700 Calibrator Replacement Battery 7.2V 2500mAh - is backordered and will ship as soon as it is back in stock.
Let customers speak for us
Send Your Battery Photo
Expert Technician Help
Snap a photo or video of your battery and send it to us. We'll identify the exact replacement—fast and hassle-free. Our team has helped thousands of customers find the right battery quickly and easily.
POST YOUR BATTERY IMAGE
Product & Solutions Expert
✉ sales@batteryweb.com
Battery Care Tips
Battery Care Tips
🔹 Getting Started
Charge your new battery fully before you use it for the first time. Over the next few charge cycles, run your device down to around 20% before you recharge—this helps the battery perform its best. After that, charge whenever you need to.
🔹 Keep It Healthy
Avoid letting your battery completely drain or staying plugged in constantly. Both extremes wear it out faster. Store the battery in a cool, dry place when you're not using it, since heat damages batteries quickly.
Delivery and Shipping
Delivery and Shipping
🔹 Most orders ship the next day, and we use FedEx, UPS, Purolator and other carriers to get them to you. Lithium batteries have to ship by ground only, not air or USPS. Make sure your address is right before you order, because if we have to send it back, you pay for shipping again.
Disclaimer
Disclaimer
⚠️ Disclaimer: All product names, trademarks, and registered trademarks belong to their respective owners.
🔹 We use these names, brands, or model numbers only for identification and compatibility purposes.
Fluke 700 Calibrator Replacement Battery 7.2V 2500mAh - is backordered and will ship as soon as it is back in stock.
Voltage
7.2V
Amp
2500mAh
Fluke 700 / 740 / 744 Calibrator Series — 7.2V Ni-MH Replacement Battery (BP7235)
This is a 7.2V 2500mAh Ni-MH replacement battery for the Fluke 700, 740, and 744 Calibrator series, along with the DSP-100 and compatible models. It slots into the battery compartment and powers the calibrator during portable field use and bench work. The OEM part numbers this pack replaces are 116-066, 668225, and BP7235.
- Calibrator series compatibility: The 700, 740, and 744 Calibrators share the same battery bay dimensions, 7.2V rail, and connector pinout. One pack covers the full lineup without modification.
- Bench tested on actual hardware: We cycled this pack through charge and discharge on calibrator-class load profiles. The BMS handled the initialisation current spike at instrument power-up without tripping, and held voltage steady under sustained sensor drive.
- Post-install calibration cycle: After fitting this pack, run a full calibration cycle through the instrument menu before field deployment. The Fluke 700 series maps battery state during that cycle — skipping it causes the instrument to throw premature low-battery warnings on the first measurement session.
BMS lockout after the calibrator sat unused in a carry case for months
Ni-MH cells self-discharge over storage. If the pack voltage drops below roughly 5.4V — the BMS sleep threshold — the charger sees the pack as absent and refuses to begin a charge cycle. The instrument also will not power on. To recover it, apply a trickle charge at 100–200mA directly to the pack terminals for 10–15 minutes to push voltage above 6V, then transfer to the standard dock. Once the BMS re-initialises, a normal charge cycle completes without issue.
Readings resetting or logging gaps during a sustained measurement session
This happens when sustained sensor load — multiple loops active, output sourcing running — pulls enough current to cause a brief voltage sag that the instrument interprets as a power fault. It is not a capacity problem; it is a cell impedance issue in aged or cold packs. A freshly charged pack at room temperature handles the combined draw without dropout. If sags persist after a full charge, the pack has degraded beyond recovery and needs replacement.
Compatible Models
Replaces Part Numbers
Technical Specifications
Product Highlights
- Brand: Fluke
- Manufacturer: CS
- Series: Standard
- Color: Black
- Product Type: Ni-MH
- Battery Type: Ni-MH
- Warranty: 12 Months
- Bulk Orders: sales@batteryweb.com
Frequently Asked Questions
My Fluke 744 Calibrator won't recognise the new pack — the charger light doesn't come on and the instrument stays dead. What's happening?
The pack likely discharged below the BMS recovery threshold during storage, so both the charger and the instrument see no valid battery present. Apply a trickle charge at 100–200mA directly to the pack terminals for 10–15 minutes until pack voltage climbs above 6V. Then place it in the standard dock — the charger will detect it and begin a normal cycle. Once fully charged, the instrument powers on normally.
The calibrator shuts down the moment I start sourcing a 4–20mA loop signal, even though the battery indicator showed full.
A full charge indicator does not mean the pack can sustain output-sourcing current. If cell impedance is high — common in packs that have been deep-discharged or stored flat — voltage sags sharply the moment loop drive starts, and the instrument's undervoltage circuit cuts power. Charge the replacement pack fully at room temperature, then re-test. If the shutdown disappears, the old pack had degraded cells; if it persists, check that the instrument firmware is not flagging a BMS handshake error under Settings.
The battery percentage on the Fluke 700 display jumps around erratically after fitting a new pack — it reads 80%, drops to 20%, then jumps back up.
The 700 series uses a voltage-threshold indicator, not a fuel gauge. When new cells are installed, the instrument's internal reference points are still calibrated to the old, degraded pack's discharge curve. The display stabilises after one or two full charge-discharge cycles, which let the instrument re-anchor its threshold markers to the new cell voltages. Run the pack through a complete discharge in normal use, charge it fully, and the percentage reading will track consistently from that point.