Codan Argus 606 S Compatible Battery 12V 2000mAh Ni-MH
This product ships directly from our Manufacturer’s Warehouse and is usually delivered within 5 – 8 business days to your doorstep.
WECARE5
Check that your old battery model number and device model to match our description. This makes sure they work together.
We ship your order same day if you buy it before 4 PM EST.
Codan Argus 606 S Compatible Battery 12V 2000mAh Ni-MH - is backordered and will ship as soon as it is back in stock.
Let customers speak for us
Send Your Battery Photo
Expert Technician Help
Snap a photo or video of your battery and send it to us. We'll identify the exact replacement—fast and hassle-free. Our team has helped thousands of customers find the right battery quickly and easily.
POST YOUR BATTERY IMAGE
Product & Solutions Expert
✉ sales@batteryweb.com
Battery Care Tips
Battery Care Tips
🔹 Getting Started
Charge your new battery fully before you use it for the first time. Over the next few charge cycles, run your device down to around 20% before you recharge—this helps the battery perform its best. After that, charge whenever you need to.
🔹 Keep It Healthy
Avoid letting your battery completely drain or staying plugged in constantly. Both extremes wear it out faster. Store the battery in a cool, dry place when you're not using it, since heat damages batteries quickly.
Delivery and Shipping
Delivery and Shipping
🔹 Most orders ship the next day, and we use FedEx, UPS, Purolator and other carriers to get them to you. Lithium batteries have to ship by ground only, not air or USPS. Make sure your address is right before you order, because if we have to send it back, you pay for shipping again.
Disclaimer
Disclaimer
⚠️ Disclaimer: All product names, trademarks, and registered trademarks belong to their respective owners.
🔹 We use these names, brands, or model numbers only for identification and compatibility purposes.
Codan Argus 606 S Compatible Battery 12V 2000mAh Ni-MH - is backordered and will ship as soon as it is back in stock.
Voltage
12V
Amp
2000mAh
Codan Argus 606 S Syringe Pump — 12V Ni-MH Replacement Battery (110840-O)
This is a 12V 2000mAh Ni-MH replacement battery for the Codan Argus 606 S Syringe Pump and Argus infusion pump 606. It replaces OEM part numbers 110840-O and 601260. The battery fits directly into the pump's battery bay at the same voltage and capacity as the original cell pack.
- Argus 606 S and 606 platform compatibility: Both Argus 606 variants share the same 12V battery rail, connector pinout, and BMS handshake protocol. A single cell pack covers both designations without modification.
- Bench tested on actual hardware: We cycled this battery through the Argus 606 S charge and discharge sequence. The BMS accepted the cell, completed the charge cycle without flagging a chemistry mismatch, and held voltage within the expected window across multiple discharge runs.
- Post-installation self-test protocol: After fitting this battery, let the pump complete its full power-on self-test without interruption. The Argus 606 S runs a BMS verification sequence at startup — cutting power during this window stores a false battery fault that persists until the next full reboot.
Why the Argus 606 S self-test fails after a battery swap
The Argus 606 S BMS runs a learn cycle when it first encounters a new cell. On the initial charge, the charge IC applies a conservative current limit while it profiles the new Ni-MH pack. If the pump is powered down mid-cycle or the self-test is interrupted, the BMS logs an incomplete verification and will continue to flag battery faults on subsequent boots. Run one complete charge-discharge cycle before placing the pump in clinical use — this lets the BMS complete its calibration pass and clear any pending fault flags.
Charge indicator not reaching 100% on first charge
On a fresh Ni-MH cell, the Argus 606 S charge controller holds back the final charge increment until it confirms the pack's temperature and delta-V signature match expected values. A new cell's internal resistance reads slightly higher than a conditioned pack, which causes the controller to interpret full charge as not yet reached. This is not a fault — it resolves after the first complete charge cycle once the BMS has mapped the cell's actual response curve. After that first cycle, the indicator will reach 100% and hold normally.
Compatible Models
Replaces Part Numbers
Technical Specifications
Product Highlights
- Brand: Codan
- Manufacturer: CS
- Series: Standard
- Color: Green
- Product Type: Ni-MH
- Battery Type: Ni-MH
- Warranty: 12 Months
- Bulk Orders: sales@batteryweb.com
Frequently Asked Questions
The Argus 606 S is alarming low battery right after a confirmed full charge — what's happening?
The pump's BMS compares the new cell's resting voltage against a threshold calibrated to a conditioned OEM pack. A fresh Ni-MH cell has a slightly different delta-V profile, so the BMS reads it as below threshold even when fully charged. Run one complete charge-discharge cycle — this lets the BMS map the actual capacity curve of the new cell and clear the alarm. After that cycle, the low battery alarm should not recur at full charge.
The pump will not power on at all after the replacement battery sat in storage — how do I recover it?
Ni-MH cells self-discharge during storage, and if the pack voltage drops below the Argus 606 S's BMS recovery floor, the controller blocks power-on to protect the cell from damage under load. Connect the pump to mains power and leave it on charge for a minimum of four hours before attempting to power on — this allows the charge IC to trickle-charge the cell back above the BMS recovery threshold. If the pump still does not boot after that charge period, check that the charger connection is seated and the charge indicator light is active.
The Argus 606 S shuts off unexpectedly during an infusion run — is the battery faulty?
This is most likely a load-profile stress response in the first ten cycles. The infusion pump's motor-drive and solenoid draw sharp current pulses, and a new Ni-MH cell's internal resistance is higher before it is conditioned — this causes a momentary voltage sag that the BMS reads as a depleted cell and triggers a protective shutdown. The behaviour typically resolves as the cell conditions across the first several cycles. Confirm the battery is at a full charge state (resting voltage at 13.2V or above) before each infusion run during the break-in period.