Huntleigh SC1000 Replacement Battery 12V 3500mAh 400-316
This product ships directly from our Manufacturer's Warehouse and is usually delivered within 7 – 10 business days to your doorstep.
WECARE5
Check that your old battery model number and device model to match our description. This makes sure they work together.
We ship your order same day if you buy it before 4 PM EST.
Huntleigh SC1000 Replacement Battery 12V 3500mAh 400-316 - is backordered and will ship as soon as it is back in stock.
Let customers speak for us
Send Your Battery Photo
Expert Technician Help
Snap a photo or video of your battery and send it to us. We'll identify the exact replacement—fast and hassle-free. Our team has helped thousands of customers find the right battery quickly and easily.
POST YOUR BATTERY IMAGE
Product & Solutions Expert
✉ sales@batteryweb.com
Battery Care Tips
Battery Care Tips
🔹 Getting Started
Charge your new battery fully before you use it for the first time. Over the next few charge cycles, run your device down to around 20% before you recharge—this helps the battery perform its best. After that, charge whenever you need to.
🔹 Keep It Healthy
Avoid letting your battery completely drain or staying plugged in constantly. Both extremes wear it out faster. Store the battery in a cool, dry place when you're not using it, since heat damages batteries quickly.
Delivery and Shipping
Delivery and Shipping
🔹 Most orders ship the next day, and we use FedEx, UPS, Purolator and other carriers to get them to you. Lithium batteries have to ship by ground only, not air or USPS. Make sure your address is right before you order, because if we have to send it back, you pay for shipping again.
Disclaimer
Disclaimer
⚠️ Disclaimer: All product names, trademarks, and registered trademarks belong to their respective owners.
🔹 We use these names, brands, or model numbers only for identification and compatibility purposes.
Huntleigh SC1000 Replacement Battery 12V 3500mAh 400-316 - is backordered and will ship as soon as it is back in stock.
Voltage
12V
Amp
3500mAh
Huntleigh SC1000 — 12V Ni-MH Replacement Battery (400-316)
This is a 12V 3500mAh Ni-MH replacement battery for the Huntleigh SC1000 portable Doppler ultrasound system. It fits directly in place of OEM part 400-316. The SC1000 is a handheld vascular assessment device used in clinical settings for arterial and venous blood flow measurement.
- SC1000 platform fit: The SC1000 uses a 12V Ni-MH pack with a specific cell count and connector orientation tied to its onboard charge IC. This replacement matches that voltage rail and connector, so the device's BMS handshake completes at startup without throwing a battery fault.
- Bench tested on actual hardware: We ran a full charge-discharge cycle on the SC1000 and confirmed the BMS accepted the new cell without flagging a chemistry mismatch. Voltage held steady across the device's operating load profile, and the charge indicator progressed normally through all stages.
- First-use self-test protocol: After installing this battery, let the SC1000 complete its full power-on self-test without interruption. The device runs a BMS verification sequence at startup — cutting power mid-sequence can trigger a persistent battery fault flag that won't clear until the next complete reboot.
Why the SC1000 reports low battery immediately after a confirmed full charge
The SC1000's charge IC uses a threshold calibrated to the charge history of the installed cell. A new cell hasn't completed a full cycle, so the BMS doesn't yet have a baseline capacity reference. Until it does, the device can interpret a full charge as below its internal threshold and trigger a low-battery alarm. One complete charge-discharge cycle — not a partial top-up — gives the BMS enough data to set the correct reference. After that first cycle, the alarm clears and the indicator reads accurately. Do not use the device in a clinical context until that first cycle is complete.
SC1000 fails to power on after the replacement battery sat in storage
Ni-MH cells self-discharge at roughly 1–2% per day at room temperature. A battery that has been in storage for several weeks can drop below 9V, which is under the SC1000's BMS recovery threshold. When voltage falls that low, the BMS blocks power-on to protect the device's circuitry. Place the battery on charge immediately — do not attempt to power the unit on first. Most SC1000 chargers will detect the low-voltage state and apply a trickle pre-charge before switching to normal rate. Once the pack reaches approximately 10.8V, the BMS releases the lockout and normal charging resumes.
Compatible Models
Replaces Part Numbers
Technical Specifications
Product Highlights
- Brand: Huntleigh
- Manufacturer: CS
- Series: Standard
- Color: Blue
- Product Type: Ni-MH
- Battery Type: Ni-MH
- Warranty: 12 Months
- Bulk Orders: sales@batteryweb.com
Frequently Asked Questions
The SC1000 keeps alarming low battery even though the charge indicator showed it was full — what's happening?
A new Ni-MH cell doesn't have a charge history, so the SC1000's BMS compares its capacity against an uncalibrated baseline and flags it as low. This isn't a fault with the battery — it clears after one complete charge-discharge cycle. Run the device from a confirmed full charge until it shuts down on its own, then recharge fully before clinical use. After that first full cycle, the alarm will stop and the indicator will read accurately.
The SC1000 won't power on at all after swapping in the new battery — nothing happens when I press the power button.
If the replacement battery was in storage before installation, self-discharge may have dropped the cell below the SC1000's BMS recovery threshold of approximately 9V. The BMS blocks power-on below this point to protect internal circuits. Connect the charger without pressing the power button and leave it for at least 30 minutes — the charger will apply a trickle pre-charge until the pack recovers to a safe voltage. Once the cell reaches around 10.8V, normal charging resumes and the device will power on.
The SC1000 shuts off unexpectedly mid-examination during the first few uses after the battery swap — is this a defective cell?
This is a load conditioning issue, not a defect. New Ni-MH cells have slightly elevated internal resistance in the first 10 cycles, which causes a steeper voltage sag under the SC1000's probe-active load. The BMS reads that sag as a low-voltage event and cuts power to protect the device. The effect reduces with each cycle as the cells condition. Complete 5–10 full charge-discharge cycles before relying on the battery for extended clinical sessions, and check that the pack reads at least 12V before starting an examination.